Overview:
The animal respirator is a commonly used experimental device, widely used in scientific research experiments such as basic medicine, clinical medicine, and animal for artificial respiration, respiratory management, emergency treatment of animals, respiratory therapy, etc. It has accurate control, convenient and practical, does not require a high-pressure air source and has precise tidal volume output and stable performance.
Working principle:
This type of animal respirator is a positive pressure respiration of constant volume, powered by a motor controlled by a drive circuit, which outputs air rhythmically, enters the animal's lungs through the airway, expands the lungs, and achieves the purpose of gas exchange.
Technical indicators:
1.Instrument size: 335x260x100mm,
2.Resp rate adjustment range: 1 – 200 times/min,
3.Tidal volume adjustment range: 0.1-100ml (mouse) 0.1-200ml (rabbit)
4.Inspiratory-expiratory ratio adjustment: 1:5~5:1,
5 Airway pressure waveform display range: 7Kpa,
5.Automatic pressure relief protection when airway pressure exceeds 6Kpa,
6.Ventilator interface: 6mm diameter pagoda connector, material stainless steel,
7.Output air volume calibration function: 1~80cc,
8.LCD brightness adjustment adjustable,
9.Parameter display and operation: 5-inch touch LCD display,
10.Screen resolution: 854x480,
11.Input voltage: AC 110~220V/50Hz,
12.Working environment temperature: 0 - 50℃,
13.Working environment relative humidity: <85%,
14.Use air or other gases as the gas source, no need for high-pressure gas, to use,
15.Applicable scope: suitable for small mice, large rats, guinea pigs, rabbits and other animals,
17. There are seven specifications tracheal intubation: 14G, 15G, 16G, 18G, 20G, 22G,25G.
bio-equip.cn