Profile
In 1931,the CIE(Commission Internationale De L'E'clalrage)setup a series of visible spectrum color space standards.The Basic CIE color space standardS iS CIE_XYZ.lt iS built on standard visual observerson the basis of ability to see,and it reflects the standardS scope of color by eye.Based On CIE_XYZ there are CIE_XvY,CIE_Lab,CIE Lch standard color space,etc.
At present,the CIE Lab color space iS the most commonly used in the industry.In ClE Lab color space,L value stands for color brightness and a value stands for the Green-0n-red value.and b value stands for the blue—on—yellow color value.If determine a value of color by only a group Lab,it will be not practical.However,we can determine the differences among them by comparing two groups of Lab values.
◆Technical Features:
⊙Easy operation make you use it easily
⊙Widely applied to quality control of plastic and printing industries’
⊙Display directly color difference by △E*ab、△L*a*b、CIE_L*a*b、CIE_L*C*h
⊙Standard deviation within△E*ab0.2(test condition:choose average values by 12pcs white tabula)
⊙It can be connected with computer to do the inspection by software with USB expansion interface.
◆Skills Indications
test accuracy |
Within 0.2 △E*a b |
Display |
△E*ab、CIE_Lab、△L,△a,△b、CIE_Lch |
Test swpe |
L:0—100 a:-128--127 b:-128--127 |
Test time |
About 3 seconds |
Test Interval |
2 seconds |
Test aperture |
¢8mm |
Automatic shutdown |
Automatic shutdown after 5 minutes waiting |
Light source |
C light source |
Sensor |
Correct sillicon photodiode carrays |
Power |
DC/5V(1.5A).2pcs 1.5V(AAA)batteries |
Size |
171*50*48.8mm |
Weight |
203.6g(Exclusive battery) |
Ope rated temperature scope |
0-40 ℃132-104'f);lower than 85% relative humidlty |
Standard accessories |
1 pc nlanual/softwbag/9V battery/5v external power supply/USB connecting cable/CD |
Application |
Measure any color Of smooth surface |
◆CIE color space coordinate chart ◆Connecting with computer to do the analysis
COLOR DIFFERENCE FORMULA
△E*ab=[(△L*)2+(△a*)2。+(△b*)2]1/2
△L*=Lsample一L*standard(lightness difference)
△a*=a*Lsample一a* standard(red/green diference)
△b*=b*Lsample—b standard。mmd(yellow/blue difference)
◆Workin~theory
Compa re the color difference between sample and the measured subject then
output CIE_Lab 3 groups data,and the 4 groups of △E,△L,△a,△b data after
comparision
△E stands for the totaI color difference.
△L+stands for partial white,△L-stands for partial black(cornpa ring with p rototype)
△a+stands for pa rtial red,△a—stands for partial green(comparing with p rototype)
△b+stands for pa rtial yellow,△b-stands for pa rtial blue(comparing with plrototype)
Color dif SCOpe |
Color dif.anylysis |
0.0 25AE |
very small or no;very perfect matching |
0.25—0.5△E |
small;acceptable matching |
0.0—1.0△E |
small to medium:acce-ptabIe in some areas |
1.0—2.0△E |
Medium:acceptable in some a reas |
2.0—4.0△E |
has dislance:acceptable in special applications |
4.0 △E |
ve ry big;not acceptable in most applications |
bio-equip.cn