Features:
1 Wide wavelength range, satisfying requirements of various fields
2 Five options for spectral bandwidth selection, 5nm, 4nm, 2nm, 1nm and 0.5nm, made according to customer's need and satisfying the requirements of pharmacopoeia
3 Fully automated design, realizing the simplest measurement
4 Optimized optics and large scale intergrated circuits design, light source and receiver from world famous manufacturer all add up to high performance and reliability
5 Rich measurement methods, wavelength scan, time scan, multi-wavelength determination, multi-order derivative determination, double-wavelength method and triple-wavelength method etc., meet different measurement requirements
6 Automatic 10mm 8-cell holder, changeable to automatic 5mm-50mm 4-position cell holder for more choices
7 Data output can be obtained via a printer port and a RS-232 interface (RS-485 and USB port optional)
8 Parameters and data can be saved for user's convenience
9 PC controlled measurement can be achieved for more accurate and flexible requirement
Specifications:
Wavelength Range: 190-1100nm
- Spectral Bandwidth: 2nm (5nm, 4nm, 1nm, 0.5nm optional)
- Wavelength Accuracy: ±0.3nm
- Wavelength Reproducibility: ≤0.15nm
- Optical System: Double beam
- Photometric Accuracy: ±0.3%T(0-100%T), ±0.002A(0-1A)
- Photometric Reproducibility: ≤0.15%T
- Working Mode: T, A, C, E
- Photometric Reproducibility: -0.3-3.5A
- Stray Light: ≤0.05%T(NaI, 220nm, NaNO2 340nm)
- Baseline Flatness: ±0.002A
- Stability: ≤0.001A/h( at 500nm, after warming up)
- Noise: ±0.001A( at 500nm, after warming up)
- Display: 6 inches high light blue LCD
- Detector: silicon photo-diode
- Power: AC 220V/50Hz, 110V/60Hz, 180W
- Dimensions: 630×470× 210 mm
- Weight: 26kg
bio-equip.cn