• Centralized distribution systems designed to enhance existing water purification systems such as the PURELAB Prima; storing, controlling and distributing prepurified water throughout a building via a loop
• Incorporates an array of system operations, monitors and alarms to improve the management of your loop system
• Standard CENTRA systems deliver a loop flowrate of up to 18 liters per minute. The CENTRA RDS is also available as high flow variable speed model, (HFV) capable of delivering up to 30 liters per minute
Standard CENTRA Model
|
RDS
|
Loop Flow rate
|
Up to 18 liters per minute
|
High Flow CENTRA Model
|
RDS-HFV
|
Typical Loop Flow Rate
|
Variable speeds, up to a maximum of 30 liters per minute 1
|
Reservoir Capacity
|
350 liters
|
Recirculation Pump
|
√
|
UV Photo-oxidation
|
√
|
0.2μm Filter
|
√
|
Optional Deionization
|
√
|
Product Water Specifications
The performance specifications given below are intended only as a guide.
|
Inorganics (resistivity @ 25°C)
|
From feed quality up to 18.2 MΩ-cm 3
|
TOC – typical
|
From feed quality up to <10 ppb 3
|
Bacteria – typical 2
|
<5 CFU/ml 2
|
Subject to suitable operation, maintenance and sanitization
1 Subject to the distribution loop design
2 Subject to correct operating and maintenance procedures 3 Subject to use of suitable deionization cylinder
bio-equip.cn