Product Description
This analyzer is using impedance principle (colorimetry for hemoglobin measurement) to categorize and count blood cells in blood.
Features
- 3-part differentiation of WBC; 20 parameters+3 histograms
- 2 Counting mode: whole blood and prediluted
- Throughput: 60 samples per hour
- Closed tube sampling optional
- Up to 50000 sample results (including histograms) can be stored
- Built-in thermal recorder, optional external printer
- 10.4'' TFT color LCD with touch screen
Parameters
WBC, LYM#, MON#, GRA#, LYM%, MON%, GRA%, RBC, HGB, MCHC, MCH, MCV,
RDW-CV, RDW-SD, HCT, PLT, MPV, PDW, PCT, P-LCR and Histograms for WBC,
RBC and PLT
Principles
Electrical resistance for counting WBC, RBC and PLT
Performance
Parameter Precision (CV %)
- WBC (109/L) 2.0(7.0-15.0)
- RBC (1012/L) 1.5(3.50-6.00)
- HGB (g/L) 1.5(110-180)
- MCV (fL) 0.4(80.0-110.0)
- PLT (109/L) 4(100-500)
Sample Volume
Prediluted: 20µL
Whole Blood: 9.8µL
Aperture Diameter
WBC 80µm RBC 50µm
Throughput: 60 samples per hour
Display: 10.4’’ Large Color LCD with touch screen
Screen Resolution: 800×600
Alarms: Error messages
Carryover
WBC/RBC/HGB<0.5%,PLT<1.0%
Input/Output
RS-232, USB, LAN, keyboard and mouse interface
Printout
Thermal recorder, 55mm width paper, various printout formats, external printer(optional)
Operating Environment
Temperature: 15℃~35℃
Humidity: 10%~90%
Power requirement: AC 100-240V, 50/60±1Hz
Dimension L×W×H(mm):436×363×367
Net weight: 18Kg
Contact
Mail: info@wis-medical.com
WhatsApp/Wechat: +86 16675444525
www.wis-medical.com
bio-equip.cn