Temperature Sensor
|
Range
|
MX2301 and MX2305 internal sensors: -40 to 70°C (-40 to 158°F)
MX2302 external temperature sensor: -40 to 70°C (-40 to 158°F)
MX2303 and MX2304 external sensors: -40 to 100°C (-40 to 212°F), with tip and cable immersion in fresh water up to 50°C (122°F) for one year
|
Accuracy
|
±0.25°C from -40 to 0°C (±0.45 from -40 to 32°F)
±0.2°C from 0 to 70°C (±0.36 from 32 to 158°F)
±0.25°C from 70 to 100°C (±0.45 from 158 to 212°F), MX2303 and MX2304 only
|
Resolution
|
0.04°C (0.072°F)
|
Drift
|
<0.01°C (0.018°F) per year
|
RH Sensor (MX2301, MX2302 only)
|
Range
|
0 to 100% RH, -40° to 70°C (-40° to 158°F); exposure to conditions below -20°C (-4°F) or above 95% RH may temporarily increase the maximum RH sensor error by an additional 1%
|
Accuracy
|
±2.5% from 10% to 90% (typical) to a maximum of ±3.5% including hysteresis at 25°C (77°F); below 10% RH and above 90% RH ±5% typical
|
Resolution
|
0.05%
|
Drift
|
<1% per year typical
|
Response Time: Temperature (typical, to 90% of change)
|
Without Solar Radiation Shield
|
With RS1/M-RSA Solar Radiation Shield
|
With RS3-B Solar Radiation Shield
|
MX2301 internal sensor
|
17 minutes in air moving 1 m/sec
|
24 minutes in air moving 1 m/sec
|
N/A
|
MX2302 external sensor
|
2 minutes, 30 seconds in air moving 1 m/sec
|
6 minutes in air moving 1 m/sec
|
5 minutes in air moving 1 m/sec
|
MX2303/MX2304 external sensors
|
3 minutes in air moving 1 m/sec; 20 seconds in stirred water
|
7 minutes in air moving 1 m/sec
|
4 minutes in air moving 1 m/sec
|
MX2305 internal sensor
|
17 minutes in air moving 1 m/sec
|
24 minutes in air moving 1 m/sec
|
N/A
|
Response Time: RH (typical, to 90% of change)
|
Without Solar Radiation Shield
|
With RS1/M-RSA Solar Radiation Shield
|
With RS3-B Solar Radiation Shield
|
MX2301 internal sensor
|
1 minute in air moving 1 m/sec
|
1 minute in air moving 1 m/sec
|
N/A
|
MX2302 external sensor
|
4 minutes in air moving 1 m/sec
|
4 minutes in air moving 1 m/sec
|
4 minutes in air moving 1 m/sec
|
Logger
|
Radio Power
|
1 mW (0 dBm)
|
Transmission Range
|
Approximately 30.5 m (100 ft) line-of-sight
|
Wireless Data Standard
|
Bluetooth Smart (Bluetooth Low Energy, Bluetooth 4.0)
|
Logger Operating Range
|
-40° to 70°C (-40° to 158°F)
|
Logging Rate
|
1 second to 18 hours
|
Logging Modes
|
Fixed interval (normal, statistics) or burst
|
Memory Modes
|
Wrap when full or stop when full
|
Start Modes
|
Immediate, push button, date & time, or next interval
|
Stop Modes
|
When memory full, push button, date & time, or after a set logging period
|
Time Accuracy
|
±1 minute per month 0° to 50°C (32° to 122°F)
|
Battery Type
|
2/3 AA 3.6 Volt lithium, user replaceable
|
Battery Life
|
2 years, typical with logging interval of 1 minute and Power Saving Mode disabled; 5 years, typical with logging interval of 1 minute and Power Saving Mode enabled. Faster logging intervals and statistics sampling intervals, burst logging, remaining connected with the app, excessive downloads, and paging may impact battery life.
|
Memory
|
128 KB (84,650 measurements, maximum)
|
Full Memory Download Time
|
Approximately 60 seconds; may take longer the further the device is from the logger
|
Dimensions
|
Logger housing: 10.8 x 5.08 x 2.24 cm (4.25 x 2.0 x 0.88 in.)
External temperature sensor diameter: 0.53 cm (0.21 in.)
External temperature/RH sensor diameter: 1.17 cm (0.46 in.)
External sensor cable length: 2 m (6.56 ft)
Solar radiation shield bracket: 10.8 x 8.3 cm (4.25 X 3.25 in.)
|
Weight
|
Logger: 75.5 g (2.66 oz) Solar radiation shield bracket: 20.4 g (0.72 oz)
|
Materials
|
Acetal, silicone gasket, stainless steel screws
|
Environmental Rating
|
NEMA 6 and IP67
|
The CE Marking identifies this product as complying with all relevant directives in the European Union (EU). This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of the FCC Rules.