BMR-A5U/D Autoclavable/SIP Fermentor
1. INTRODUCTION
★It is a indirect driven with high toque motor fermentor with glass vessel(SIP fermentor:top-glass, bottom-stainless steel).
★It contain a ABCS Bioprocess Controller(industry computer /operating system: windows XP) which has a large size LCD screen(12.1” at least ):
※ Measurement and control parameters: temperature , pH , DO , agitator speed , antifoam , pressure(manual) , air flowrate(manual)
※ Control mode: manual control, automatic control, sequence control
※ The controller also can display the trends, historic trends, report forms and so on.
※ The date of process will be saved on hard disk(40G) . You can copy or transmit the data to a personal computer and deal with or print on EXCEL-based.
2. SPECIFICATIONS
(1) Basic Systems
Ⅰ.Glass Vessel: Total volume:5L
Working volume:3.5L
Ⅱ.Agitation:50~1000±5rpm
Ⅲ.Temperature :(temperature of cooling media + 5℃)~50℃±0.2℃
(PID with pulse-width modulation of heater in waterbath & cooling water)
Ⅳ.pH :2.00~12.00±0.05pH
(Sensor: HAMILTON pH sensor
PID with pulse-width modulation
Two pumps each assignable to function for acid, base.)
Ⅴ.D.O. : 0~150±3% . Digital display in 0.1% increment
(Sensor: HAMILTON D.O. sensor
PID with pulse-width modulation
Via agitation module or flowrate module)
Ⅵ.Anti-Foam: PID on/off control with a pump
Ⅶ.Nutrition: on/off control with a pump
Ⅷ.Air flowrate: 0~5L/min (control by manual flowmeter or you
can choose automatic control)
Ⅸ.Pressure :0-0.3MPa(manual)
Ⅹ.Electrical Date:Power-110/220AC , 1 kW
Ⅺ.Overall Dimensions(length * width * height):600*500*600mm
Weight: about 40kg
(2) You can choose some additional functions such as:
① Measurement and control of air flowrate , pressure and level
② Measurement of O2 , CO2 exhausting
③ Add two pumps for Nutrition
④ Add electron balance systems
bio-equip.cn