Operating Range: |
0°C to 65°C standard |
Ambient Temperature: |
10°C to 40°C non-condensing |
Stability / Repeatability: |
± 0.1°C at constant load (even near ambient) |
Cooling Capacity1: |
275 Watts @ 20°C (20°C ambient) |
Noise (at 1 meter): |
< 63 dBA |
Process Fluid: |
PAO (Poly-Alpha-Olefin), Options available for Fluorinert / Galden or HFE |
Process Fluid Fittings: |
1/4" CP metal (CPM) |
Pumps: |
1, 2 or 3 lpm at 30 psig magnetically
coupled gear pump |
Tank Volume: |
900 ml stainless steel tank with level sensor |
Wetted Materials: |
Aluminum, stainless steel and polymers |
Fluid Filter: |
5 μm external filter with grounding strap |
Size (L x W x H): |
13" x 11" x 13" (32 x 28 x 32 cm) without filter |
Weight: |
34 lbs (15.4 kg) |
Power Input: |
Universal: 115-230 VAC, 50/60 Hz, 7-5 amp max |
Communications: |
Keypad or RS232 Interface |
Alarms: |
Temperature, fluid level, system or component
failure (display, RS232 and Dry Contact) |
Standards: |
TUV listed to UL, CAN/CSA and EN 61010-1,
CE 61010-1 |
Warranty: |
2 years |