African Swine Fever Virus Antibody Rapid Test Kit
(Whole Blood / Serum / Plasma / Milk)

PRINCIPLE
Quantitative or qualitative detection of the titer level of African Swine Fever Virus antibodies in samples using a double antigen sandwich method. During detection, the African swine fever antibodies in the sample binds to the colloidal gold-coated specific antigen fragment to form an antigen-antibody complex, which flows along the test paper to the other end of the NC membrane. When the complex flows to the T-zone on the membrane, the antigen immobilized on the membrane captures the complex and gradually agglutinates into a red visible T-line, while the unbound colloidal gold antibody flows through the T-zone and is captured by the secondary antibody in the C-zone and forms a visible red C-line.
Naked eyes interpretation for Qualitative Results , Quantitative Analysis with machine.
KIT CONTENTS
NO. |
NAME |
PACKING |
1. |
Rapid Test Cassette |
20 Tests |
2. |
Plastic Dropper |
20 Tubes |
3. |
Technical Manual |
1 Set |
STORAGE AND STABILITY
1. Store in a light-proof, dry environment (2-30℃).
2. DO NOT FREEZE or use beyond the expiration date.
3. The shelf life is 18 months.
QUALITATIVE ANALYSIS
1. Sample dilution to be tested: Pipette 35ml (or 1 drop) of serum sample, add to the sample dilution tube, gently invert up and down 10 times and mix thoroughly; The sample can be used for testing after dilution and can be stored for 1 day at room temperature.
2. Test card restored to room temperature: The test should be carried out at 20-30℃. If the test card is stored in a refrigerator (2-8℃), equilibrate it in a room temperature environment for 30 min before ripping the packaging inside the aluminum foil bag to avoid abnormal measurement results due to moisture absorption.
3. Test card number: After preparing the sample to be tested, remove the required test card and mark the sample number on the card.
4. Dropwise sample reaction for 20 min: Aspirate 70 ml of the serum sample to be tested, drop it into the sampling hole and leave it horizontally for 20 min.
5. Results: If the C-line of the test card shows color, it indicates that the test is normal and the experiment is effective; otherwise, the experiment is invalid and it is recommended to retest.