Advantage:
antigen detection can directly detect whether the human body contains, which can be detected at the early stage of virus infection. It is fast, accurate, and requires less equipment and personnel. Using double antibody sandwich method and using two antigen specific antibodies to identify and combine different epitopes of a target antigen, it can greatly reduce the number of cross reactions. So as to improve its specificity.
Sample collection:
Sample testing:
The results of judgment:
Positive: T-line and C-line present within 15-20 minutes after sample addition.
Negative: only C-line, no T-line, occurs within 15-20 minutes after dropping the sample.
Invalid: As long as there is no line C, this test card is invalid. Please use another test card to retest.
The results are summarized in the following table. The RT-PCR cycle threshold (Ct) is the relevant signal value. Lower Ct value indicate higher viral load. The sensitivity was calculated for the different Ct value range (Ct value≤33 and Ct value≤37).
Antigen |
RT-PCR (Ct value≤33) |
Total |
Positive |
Negative |
Positive |
152 |
3 |
155 |
Negative |
5 |
601 |
606 |
Total |
157 |
604 |
761 |
Sensitivity (Ct≤33): 96.8% (152/157), (95%CI: 92.8%~98.6%)
Specificity: 99.5% (601/604), (95%CI: 98.6%~99.7%)
Accuracy: 98.9%((152+601)/761), (95%CI: 98.2%~99.5%)
Technical Parameters:
Product Name |
Main composition |
Testing time |
|
|
|
|
|
Antigen Test Kit(Colloidal Gold) |
Test card,
Sample extraction buffer,
Nasal swab or Throat swab |
15-20min |
|
-- Competitive price and bulk stock could be guaranteed .