Preparation of dac150 chromatographic system
system function
With 150 mm dynamic axial compression column.
Two 1000ml / min 10MPa infusion pumps and mixers were used to achieve two-dimensional high-pressure gradient elution.
Two 1000ml / min 10MPa infusion pumps were used as sample injection pumps to realize automatic sampling under the control of workstation software.
Variable wavelength UV detector can be used to detect four wavelengths at the same time. Four different wavelengths can be obtained by one sampling.
The chromatographic workstation meets the requirements of fda21 CFR Part 11.
Automatic collection, elution and separation of samples and automatic collection of components are realized under the control of chromatographic workstation.
Optional automatic component collector.
Specifications and detailed parameters
pump
Speed range
0.1-1000 ml / min
Structure of infusion pump
Double plunger parallel type;
Flow accuracy
< ± 1%
Velocity reproducibility
< ± 0.5%
Pressure range
0-10 MPa (1ml / min, secondary water, room temperature)
Gradient precision
1%
detector
Wavelength range
190-400nm deuterium lamp, 190-740nm deuterium lamp and halogen tungsten lamp
Cut off the filter
370 nm
light source
Deuterium lamp, halogen tungsten lamp (optional)
bandwidth
8 nm
Wavelength accuracy
± 1 nm
noise
two × 10-5au (254nm @ 1ml / min @ methanol)
drift
fifteen × 10-5au / h (254nm @ 1ml / min @ methanol)
measuring range
0 3AU
workstation software
Meet the requirements of fda21 CFR Part 11 and GLP / GMP
Control: automatic injection, elution separation, automatic component collection
Data processing: the workstation uses database storage to ensure the integrity, authenticity and security of data; Support multi-level authority management (administrator, supervisor, user, etc.).
Dac150 preparative chromatography system is equipped with a 150 mm dynamic axial compression column medium-sized amplification system. The column preparation and column installation system adopts integrated design. The user can install the column by himself. The packing always keeps the axial dynamic pressure in the column, and can keep the stability and efficiency of the column bed for a long time. After the preparation column is installed, it can be used online without disassembly. The single sample size is about 10g, which is suitable for customers whose annual output is less than 50kg.
bio-equip.cn