Home >Products> Medical Lab Instruments >Microplate Readers / Plates Washers> Rayto Elisa reader Microplate Reader RT-2100C
Rayto Elisa reader Microplate Reader RT-2100C
Rayto Elisa reader Microplate Reader RT-2100C
Origin of place China
Model RT-2100C
Supplier Wis-Medical
Price 950USD/unit
Hits 3984
Updated 3/11/2024
  • Product Detail
  • Company Profile

RT-2100C  support COVID-19 Elisa Reagent Test

  • Classical model and stable performance
  • Fixed time, End Point methods
  • Large LCD display with touch screen
  • 4 fixed filters and 4 more filters optional
  • 8-channel optical fiber system
  • 5 seconds reading for whole plate
  • Auto self-check when power on
  • Powerful QC function

Program setting

You must set the parameters before you perform a new program.

In the main menu, click program,enter the program setting window,all the already set program will be shown in it. If you want to add a new one, click new, the new program window will popup, here are the parameter (take HbsAg for example):
Program:Input English program name,e.g. HbsAg.(necessary)
Full name:Input the full name of the program,e.g.” hepatitis B surface antigen”(you can ignore it)
Reagent:Input the reagent manufacturer name,you can ignore it.
First wavelength:450 ,According to the reagent instruction   
Second wavelength:Choose 630 (recommendatory)
Calibration mode:According to the reagent instruction e.g.Cut-Off(usually used in qualitative program)
Double sample:Each sample do two tests and take the average for result.(usually not choose)           Blank≤0.05 (use default)Click  》》 to enter the next setting window.
Cuf-Off Expressions:According to the reagent instructions. E.g.:2.100 * NC +0.000 * PC +0.000
Range of Negative
:According to the reagent instructions.E.g.0.050~3.500
Range of Positive
:According to the reagent instructions.E.g.0.000~3.500
Double of Control
:Set to use double control e.g. two negative two positive, usually not used.click  》》 to enter the next setting window.
Qualitative:According to its judge method,this program use“Positive threshold value”Negative ≤ 1.0 S/Co        Positive > 1.0 S/Co    (the two parameters usually not changed)
Normal:Input normal range, this program 0.000~1.000  
Notes:Positive threshold range 0~1negative threshold range 1~9999

click Finished to finish the setting,you can also set other programs follow these instructions.

 

Contact

Mail: info@wis-medical.com

WhatsApp/Wechat: +86 16675444525

www.wis-medical.com

bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号