Yaxin-1105 Photosynthesis & Fluorescence System
The instrument can be used to calculate the rate of photosynthesis (respiration) and transpiration, stomata conductance, internal CO2 concentration of leaf, as well as obtain the efficiency of quantum of light or water efficiency etc..
I. Instrument Functions
1.Measurement:CO2 concentration, humidity, temperature, flow, leaf temperature, PAR (light intensity), atmospheric pressure. Parameters which indicate Chlorophyll fluorescence intensity such as Fo, Fj, Fi, Fp.
2.Calculation: Photosynthetic rate, transpiration rate, stomatal conductance, intercellular CO2 concentration
Fluorescence parameters: Fv, Fv/Fm, Vj, Mo, PIABS, Area, tFm, etc.
3.Control Function: Control the flow automatically, control the PAR automatically, control the measuring time, open and close style system optional, manual or automatic measurement, control the action light
4.Graphic Function
Photosynthetic curve(photosynthesis-light intensity curve)
Display OJIP-Chlorophyll Fluorescence Induction Kinetics Curves
5.Interface Function
Display parameters, measuring values, system information in real-time
6.Software
Embedded flash memory chips
Interface:32G SD card, USB2.0
Automatic firmware update
Position machine software free of installation
7.Power Supply
Li-ion rechargeable
Artificial light source, independent power supply enable continuous work
8.Module Function
Photosynthetic module
Fluorescence module
Photosynthetic-light module
Ⅱ. Technical Parameters
1.CO2 Sensor
Type: The diffuse infrared analyzer; Range:0~1500ppm; Resolution:0.1ppm; Accuracy: average noise≤0.4ppm; maximum noise≤0.8ppm (CO2 constant); Response time:15s
2.Light Quantum Sensor
Type: silicon photocell with a fixed optical filter; Range:0~2500μmolE/m2s; Spectrum:400~700nm; Resolution:1μmolE/m2s; Error: ±5%
3.Leaf Temperature Sensor
Type: T thermocouple; Range:0.0~50.0℃; Resolution:0.1℃; Error:±0.3℃
4.Atmospheric Temperature Sensor
Type: Thermistor; Range:0~50℃; Resolution:0.1℃; Error:±0.3℃
5.Flow Meter
Type: micro electron flow meter; Range:0.3~1.0L/min; Resolution:0.01L/min; Error:5%
6.Humidity Sensor
Type: Sensirion digital; Range:0~90%RH; Resolution:0.1%; Error:<±1.8%RH
7.Host
Display:240×128 dot-matrix LCD display; Memory capacity:32G SD card;8MB flash chip; Interface: USB2.0; Power supply:7.2V 10Ah, Li-ion rechargeable,7 to 9 hours battery life of one charge; Size:28.0×24.5×16.0cm; Weight:5.0kg; Others: firmware update
8.Fluorescent Probe
Type: YX handheld
Light source: blue LED,470nm
Range of light intensity:3500μmolE/m2s
Measuring time:1s~10s, adjustable
Fastest sampling speed:5μs
Weight:200g
Size:2.5×9×5cm
9.Optic-controlled Module
Type: YX-LA
Light type: red LED, blue LED
Spectrum:630nm,470nm
Type: made up of red and blue LED
Range:0~3500μmolE/m2s
Light-emitting area:3.0×7.5cm
Size of lamp base:10×3.5×5.5 cm
Size of power supply:12×8×4cm
Power supply:7.2V 10Ah, Li-ion rechargeable
Light intensity control: automatic
10.Leaf Chamber (choose any one)
Ⅰ:(25×25mm)
Ⅱ:(55×20mm)
Ⅲ:(65×10mm)
YXLC-4(85×45mm) moss leaf chamber
YXLC-5 (130×ø90mm) areatus leaf chamber
11.Operating Conditions
Temperature:0~50℃; Humidity:0~90%RH (non-condensate)
bio-equip.cn
For over 10 years of professional experience in the field of design and manufacture instruments of plant physiology, Beijing Yaxinlily Science and Technology Co.,Ltd. (Yaxin) is the pioneer in life research instrument designer in China. Its most portable and effective instruments have been used in many labs and well known by the scientist and exporters in China.
Nowadays, Protable Photosynthesis System, leaf Meter, Bio-O2 Meter, Chl-fluorescence Analysis, Prometer (Stomata Resistance Meter) are our powerful products. We focus on the need of the customer and feedback from them.
All our products are portable, easy for use and with long power time for work in the field.
Meanwhile, Beijing Yaxinlily Science and Technology Company provide the more effective and closer service to our customers.