General parameters
|
Item
|
Specifications
|
Power supply
|
AC220V,50/60Hz(Standard) 1320VA; 110V 60Hz(Optional)
|
Size(L*W*H)
|
950*645*(1250-11450) mm
|
Maximum heater output power(220VAC)
|
Air chamber heater:330W; Water tank heater:155W
|
Auxiliary network power output
|
AC220V/50Hz, both maximum permissible currents are 220V 1.5A
|
Heater indication
|
0~100%, 10% increasing amount
|
Temperature control mode
|
Bed temperature mode; skin temperature mode
|
Bed temperature control range
|
25.0~37.0℃; 37.1~39.0℃(Span mode)
|
Skin temperature control range
|
34.0~37.0℃;37.1~38.0℃(Span mode)
|
Skin temperature sensor display range
|
5.0~65.0℃
|
Heating time
|
≤ 50min
|
Temperature variability
|
≤ 0.5℃
|
Temperature uniformity (mattress in horizontal position)
|
≤ 0.8℃
|
Temperature uniformity (mattress in tilted position)
|
≤1.0℃
|
Difference between average incubator temperature and control temperature
|
≤1.5℃
|
Skin temperature sensor accuracy
|
±0.3℃
|
Humidity control
|
The running time of the tank after it is filled with distilled water
|
Humidity below 70%RH, at least 24h
|
Water tank capacity
|
1500ml
|
Humidity display range
|
0%RH~99%RH
|
Humidity control range
|
CAH%RH~90%RH(1% per adjustment)
|
Humidity control accuracy
|
±10%
|
Oxygen concentration control
|
Oxygen concentration display range
|
0%~99%
|
Oxygen concentration display resolution
|
1%
|
Oxygen concentration control accuracy
|
±5% oxygen volume concentration
|
Oxygen concentration setting range
|
21% ~ 60% (1% per adjustment)
|
Oxygen sensor service life
|
The maximum use time is 10000h at 100% oxygen concentration
|
Transport storage
|
Temperature
|
-20℃~+55℃
|
Atmospheric pressure range
|
500hPa~1060hPa
|
Relative humidity
|
≤93%RH
|
Packing size/ Gross Weight
|
Base packing size: 1200*800*870mm 50Kg
Body packing size: 1100*710*855mm 60Kg
|
The work environment
|
Relative humidity
|
30%RH~75%RH
|
Environment temperature
|
+20℃~+30℃
|
Pressure range
|
700hPa~1060hPa
|
Service altitude
|
≤2000m
|
Overvoltage category
|
Ⅱ
|
Pollution level
|
2
|
Ambient air velocity
|
≤ 0.3m/s
|
Other indicators
|
Ambient noise in the infant compartment of the incubator
|
No more than 55dB(A) at stable temperature
|
Carbon dioxide concentration in the infant compartment
|
A mixture of 4% carbon dioxide and air was fed into an 8mm diameter tube in the vertical direction from the mat to the top of the box at 750mL /min to a point 10cm above the mattress center and 10cm above the center. The concentration was less than 0 5% after stabilization
|
The velocity of air flow over the baby mattress
|
≤0.35m/s
|
General
alarm
|
Power failure alarm, fan alarm, sensor alarm, deviation alarm, over temperature alarm, water tank placement error alarm, water shortage alarm
|
System alarm
|
Mcu1 and Mcu2 communication error alarm, Mcu2 and Mcu3 communication error alarm, analog-to-digital conversion chip 2544 failure alarm, analog-to-digital conversion chip 2548 failure alarm, over temperature alarm, temperature heating system failure, humidity heating system failure
|