Medlab—U/4C501 Biologic Signal Collection systems
Newly added and main improving points
①Adopt high resistance(10E12Ω , 3pF), low nose. low drift 8i high function’s amplifier fuse to be ready, so it has such advantages as strong anti—interference, good linear degree and high measuring accuracy.
②Adopt USB 2.0 structure to transmit data, so its velocity is more high, and guarantees data’s continuous collection.
③Add 9 stage Besil low-pass wave filter newly, bi close to ideal condition
④Add 50Hz hardware wave frame circuit and datum tracing circuit newly
⑤Enlarge inputting electrical level’s scope to ±1.2v ,it’s very important to this condition that there is higher polarization voltage though the signal is very weak.
⑥Magnifying times is from 1,2,5,to 5000 so it can adapt to many kinds of signals.
⑦When outputting voltage is to 100v outputting current is to 100ma from electrical stimulator, it will separate light from electricity and be safe.
Amplifier technical index
No | Item | Parameter''s demand | 1 | Channel''s number, | Four—channel amplifier has the same function adjustable parameter control and part technology | 2 | Inputting mode. | Tow sends balance inputs differentially | 3 | Inputting resistance | 100M | 4 | Gain transverter | 1, 2, 5, 10, 20, 50, 100, 200, 500, 1000, 2000, 5000 ,10000, 20000, 50000 | 5 | Gain accuracy | 0.5% (FS) ( an hour''s preliminary heating later) | 6 | Gain accuracy | 0.05%(FS) (an hour''s preliminary heating later) | 7 | Gain stability | 0.1 percent of full degree | 8 | Linear degree | <1% | 9 | Distortion factor | DC—30KHZC+0.5db—3db | 10 | Nose | <3 MVRMS( Inputting short circuit, at maximum gain 30k wide band, it will convert to in putting end) <2uvRMS(Inputting short circuit, at maximum gain 30k wide band, it will convert to outputting end ) | 11 | Common mode''s restrain | >100db | 12 | Time drift | <3MV/H (Inputting short circuit, at maximum gain, it will convert to inputting end) | 13 | Temperature drift | <3mv/h(Inputting short circuit, at maximum gain, it will convert to outputting end ) | 14 | Upper limit frequency | When voltage is DC—30kHz, its transverter is 10, you also can input upper limit frequency freely | 15 | Wave filter character | 9-stage Besil low-pass wave filter | 16 | Time constant | DC 5s,2s,1s,0.2s,0.002s, 0.001s, 0.005s, adds up to 9 transverters altogether, you can switch freely | 17 | 50 Hz wave frame | Have but attenuate 32 db | 18 | Datum tracing | Have and overcome datum drift | 19 | Monitor function | Monitor four channels freely | 20 | Note drop function | Four channel all can note including drops, drop velocity |
Stimulator parameter
No | Item | Parameter''s demand | 1 | Range | 0-100v''s distance 13/mv | 2 | Outputting current | 100mm | 3 | Wave width | 0.1-600ms, program adjustable distance is 0.1ms | 4 | Delay | 1-250ms, program adjustable distance is 0.1ms | 5 | Operating mode | Alone stimulation, crossing stimulation main period stimulation, automatic intermission stimulation, automatic range adjusting stimulation automatic wave width adjusting stimulation, automatic frequency adjusting stimulation | 6 | Stimulation pulse value | Free | 7 | Stimulation series length | free | 8 | Stimulation series intermission | Free | Operating table | 1 | Backing system | Win2000/Winxp | 2 | Data transmission | Full velocity USB structure. |