Bante900 multi-parameter meter is capable of measuring up to 9
water quality parameters, including pH, ORP, Ion, Conductivity
TDS, Salinity, Resistivity, Dissolved Oxygen, etc.
General Features:
Automatic Temperature Compensation ensures high accuracy
of measurements.
Calibration due reminder helps you to calibrate meter regularly.
Automatic endpoint function freezes stable measured values.
Help message as a operational guide to helps you understand
how to use meter.
Setup menu lets user customize the calibration points, stability
conditions, date and time to meet measurement requirements.
Expanded memory stores and recalls up to 500 readings.
Built-in real-time clock stamps stored data to meets GLP.
USB communication interface.
|
|
pH Mode:
During the pH mode, Up to 5 points push-button calibration with auto-
buffer recognition.
Selectable pH buffer standards and resolution (0.1, 0.01, 0.001pH).
Automatically calculates and displays electrode slope that helps you
decide whether to replace pH electrode.
ORP Mode:
Single point offset calibration allows user adjusts displayed value to
a known standard.
Absolute or Relative mV measurements provides accurate readings.
Ion Mode:
Standard BNC connector can connects many Ion Selective Electrode.
2 to 5 points push-button calibration from low to high concentrations.
Selectable multiple concentration units, including ppm, mg/L, mol/L.
Direct readout ion concentration simplifies measurement process.
mV measurement allows user to check performance of ion electrode.
Conductivity Mode:
Up to 5-point push-button calibration allows user to uses customized
calibration solutions.
Adjustable TDS factors, cell constants, temperature coefficients and
normalization temperature (20°C or 25°C).
Selectable linear or pure water compensation modes.
In salinity measurement modes, selectable practical salinity (PSU) or
seawater measurement standard (ppt).

Dissolved Oxygen Mode:
1 or 2 points auto-calibration (100% in saturated air or air-saturated
water, 0% in zero oxygen solution).
Manual salinity and barometric pressure compensation.
Selectable readout in ppm, mg/L or % saturation.
|
|
|
Specifications: |
|
|
|
|
Model |
|
Bante900 |
|
|
|
pH Range |
|
-2.000~20.000pH |
|
|
pH Accuracy |
|
±0.002pH |
|
|
Resolution |
|
0.001, 0.01, 0.1pH, Selectable |
Calibration Points |
|
Up to 5 points |
Calibration Solutions |
|
USA(pH1.68/4.01/7.00/10.01/12.45), NIST(pH1.68/4.01/6.86/9.18/12.45), DIN (pH1.09/4.65/6.79/9.23/ 12.75) |
Ion Concentration Range |
|
0.001~19999ppm, mg/L, mol/L (Depending on measuring range of Ion Selective Electrode) |
Ion Concentration Accuracy |
|
±0.5% Full Scale (Monovalent), ±1% Full Scale (Divalent) |
Calibration Points |
|
2 to 5 points |
Calibration Solutions |
|
0.001, 0.01, 0.1, 1, 10, 100, 1000, 10000ppm |
mV Range |
|
-1999.9~1999.9mV |
mV Accuracy |
|
±0.2mV |
Resolution |
|
0.1, 1mV, Selectable |
Conductivity Range |
|
0~20mS/cm (Max.200mS/cm) |
Conductivity Accuracy |
|
±0.5% Full Scale |
Calibration Points |
|
Up to 5 points |
Calibration Solutions |
|
10uS/cm, 84uS/cm, 1413uS/cm, 12.88mS/cm, 111.8mS/cm |
TDS Range |
|
0~10ppt (Max. 20ppt, Depending on factor setting) |
TDS Accuracy |
|
±1% Full Scale |
TDS Factor |
|
0.1~1.0 (Default 0.5) |
Salinity Range |
|
0~10ppt (Max.80ppt) |
Salinity Accuracy |
|
±1% Full Scale |
Salinity Measurement Modes |
|
Practical Salinity (PSU) or Seawater Measurement Standard (ppt) |
Resistivity Range |
|
0~100MΩ |
Resistivity Accuracy |
|
±1% Full Scale |
Resolution |
|
0.001, 0.01, 0.1, 1, Automatic |
Temperature Coefficient |
|
0.0~10.0% per °C |
Compensation Modes |
|
Linear or Pure Water |
Cell Constant |
|
K=0.1, 1, 10, Selectable |
Normalization Temperature |
|
20°C or 25°C |
Dissolved Oxygen Range |
|
0.00~20.00mg/L or ppm |
Resolution |
|
0.01mg/L |
Dissolved Oxygen Accuracy |
|
±0.2mg/L |
% Saturation of Oxygen |
|
0.0~200.0% |
Accuracy |
|
±2.0% |
Salinity Correction |
|
0~50g/L |
Pressure Correction |
|
60.0~112.5kPa, 450~850mmHg |
Temperature Range |
|
0~105°C, 32~221°F |
Temperature Accuracy |
|
±0.5°C, ±0.9°F |
Temperature Calibration |
|
Manual |
Temperature Compensation |
|
0~100°C, 32~212°F, Automatic or Manual; DO Mode: 0~50°C, 32~122°F, Automatic |
Auto-Off |
|
Manual or Automatic (10, 20, 30 minutes after last key pressed, Selectable) |
Hold Function |
|
Manual or Automatic endpoint |
Stability Conditions |
|
Low or High, Selectable |
Calibration due reminder |
|
1 to 30 days or Off |
Reset Function |
|
Yes |
Memory |
|
500 data sets with Date/Time stamping |
Output |
|
USB Communication Interface |
Connector |
|
BNC and 6-pin connector |
LCD Display |
|
Custom dual-display LCD (130×110mm) |
Power Requirements |
|
DC5V, using AC adapters, 220VAC/50Hz |
Dimensions |
|
210(L)×188(W)×60(H)mm |
Weight |
|
1.2kg |
bio-equip.cn
Tsingtao Toky Insturments Co.,Ltd is one of hi-tech factory specialized in the production of instruments and meters,all of our Products are sold to europe, japan, usa, the middle east countries and regions.