• Introduction
The Double Window Constant Low Temperature Bath is a high precision thermostatic bath which combines the microprocessor and the PID control technique. With a circulatory pump to carry out the agitation, the bath produces little noise and is pollution free. The two large windows at the front and back make it especially suitable when the working process needs to observed. It is equipped with low liquid protection, temperature out of control and electricity leakage protection.
This series of products are widely used as temperature controlled water baths in industries such as petroleum, chemical, medicine, biochemistry, metrology and other light industries and scientific research etc.
• Features
Windows at the front and back for easy observation of the process
Clear LED digital display
Thin-film panel for easy operation
Low liquid level protection
Temperature out of control protection
Protection and alarm sensors
Low liquid level protection
Over current and over pressure protection
Sensor abnormality alarm and protection
Electricity leakage protection
ModelParameter |
N-20W |
N-40W |
N-65W |
Environment temperature |
5 ~ 40 ℃ |
Temperature control range |
Room temperature+5 ~ 80 ℃ |
Temperature control accuracy |
± 0.02 ℃ |
Temperature setting accuracy |
0.1 ℃ |
Temperature display accuracy |
0.01 ℃ |
Heater |
1800W |
Main power1800W
Auxiliary power 1000W |
Main power 2500W
Auxiliary power 1800W |
Circulation pump capacity |
Jet type
Max.flow12L/min
Max. head 2.1m |
Jet type
Max.flow 19L/min
Max. head 3.4m |
Jet type
Max.flow 31L/min Max. head 4.3m |
Electricity leakage for action of protection system |
30mA |
Insulation resistor |
> 1.5M Ω |
Internal dimension
(L×W×H)(mm) |
300 × 250 × 300 |
400 × 350 × 300 |
500 × 400 × 350 |
Dimension
(L×W×H)(mm) |
490 × 310 × 410 |
590 × 410 × 410 |
690 × 460 × 460 |
Window dimension(L×H)(mm) |
240 × 215 |
340 × 215 |
440 × 265 |
Input A.C. Voltage |
AC220 ± 22V 50 ± 1Hz |
Input power |
1900W 10A |
2900W 14A |
4400W 20A |
Net weight |
25kg |
35kg |
40kg |
* Measuring condition for temperature control accuracy(±0.02℃): 1.Measuring temperature:40℃ 2.Environment temperature:25℃
bio-equip.cn
As a Japanese wholly owned enterprise, BIOER TECHNOLOGY CO., LTD. is the manufacturing base of the Bio Flux Corporation whose headquarter lies in Tokyo, Japan. BIOER is committed to researching and manufacturing of hi-tech products for life science applications. Years of experience and technical know-how have been developed into three product lines: life science instrumentation, related reagents and disposables and general laboratory instrumentation,such as Real-time PCR systems, thermal cyclers, cooling & heating blocks,Nucleic Acid Purification Systems, HBV, HCV, HIV, GMO, nucleic acid purification kits and all kinds of constant temperature baths. These products are recognized in both the overseas and domestic markets and have won praise from customers all over the world, including: Hong Kong, Taiwan, Korea, Japan, India, Malaysia & USA.
BIOER moved to its new state of the art facility in September 2003 that occupies a floor space of 16,000 square meters . BIOER uses the latest computer technology throughout its production process and strictly follows international ISO 13485 standards aiming at creating a first-class brand and products with first-class after-sales services for its clients.