Login    Join free Add to favorites    中文
Home >Products> Cryostat | Freezers | Thermostat >Low Temperature Freezers> Plasma Freezer 237L
Plasma Freezer 237L
Origin of place China
Model DXJ-4
Supplier ZHENGZHOU FEILONG MEDICAL EQUIPMENT CO.,LTD
Price
Hits 1440
Updated 5/31/2010
  • Product Detail
  • Company Profile
Commodity Name:Blood Bank Refrigerator
Usage:Apply for blood plasma freeze and stock in blood bank stations,hospitals or labs.
Maine Advantages:
1,PID temperature controlled technology,precise and steady .
2,Blood plasma can be frozen within one hour,high efficiency and speed.
3,Pogrammable large open surface color touch screen.
4,Temperature can be changed into -30 degree for storage automatically after the frozen finished.
5,100 days of historical data for looking up and download .
6,USB interface for download memory date.
7,Special designed plasma special clip,increase heat transfer area,frozen fast and evenly.
8,Initiate up to 100 degree of disinfection function to avoid the infectant from the mucedine and other pathogenic microorganism .
9,Introduced TECUMSEH brand low noise compressor unit from France, low power dissipation and noise.
Technical Date:
Controlled temperature range: -70 ~ +100 celsius
Temperature fluctuation range: +/-0.5 celsius
Temperature evenness: less than 2 celsius
Maximum circular freezing volume: 14400ml(200ml x 72bags)
Frozen goods center temperature: less than 30 celsius
Special plasma clip: 12pieces
Freezer Power: less than 7 kilowatt
Box weight:: 350kgs
Source of voltage : 380V +/- 10% ,50HZ
Exterior size: D950 * W940 * H1760mm
Working room size: D450 * W500 * H1050mm
Freezer style: air swept twin-stage
Temperature reduction of empty working room in the box:-55 celsius less than 20 minutes;-70 celsius less than 30 minutes.
Freezing process:keep the box empty and reduced the temperature to -55 celsius,put in the plasma and the temperature should not surpass -30celsius.After 30-45 minutes later or when the temperature of the box reduced to -55 celsius,the center temperature of the plasma is -30 celsius.
bio-equip.cn
Request Infomation

* Name:
Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request infomation:
yes no
Request Quotation:
yes no
* Message:
Validated Code:
refresh
I agree to share my inquiry to the other matching suppliers.



Copyright(C) 2006-2025 Bio-Equip    E-mail:web@bio-equip.cn   沪ICP备06040519号