Main Specification
|
Wavelength range
|
190-900nm
|
Wavelength accuracy
|
±0.5nm
|
Resolution
|
Two spectral lines of Mn at 279.5nm and
279.8nm can be separated with the spectral
bandwidth of 0.2nm and valley-peak energy
ratio less than 30%.
|
Baseline stability
|
0.005A/30min
|
Background correction
|
The D2 lamp background correction
capability at 1A is better than 30 times.
|
Light Source System
|
2 lamps are powered simultaneously (one preheating)
|
Lamp current adjustment range: 0~20mA,
|
Lamp power supply
mode
|
Powered by 400Hz square pulse
|
Optical System
|
Monochomator
|
Single beam, Czerny-Turner design grating
monochromator
|
Grating
|
1800 l/mm
|
Focal length
|
277mm
|
Blazed Wavelength
|
250nm
|
Spectral Bandwidth
|
0.1nm, 0.2nm, 0.4nm, 1.2nm 4 steps
|
Adjustment
|
Manual adjustment for wavelength and slit
|
Flame Atomizer
|
Burner
|
10cm single slot all-titanium burner
|
Spray chamber
|
Corrosion resistant all-plastic spray chamber.
|
Nebulizer
|
High efficiency glass nebulizer with metal
sleeve, sucking up rate: 6-7mL/min
|
Position adjustment
|
Manual adjusting mechanism for vertical,
horizontal positions and the rotation angle of
he burner
|
Gas line protection
|
Fuel gas leakage alarm
|
Detection and Data
Processing System
|
Detector
|
R928 Photomultiplier with high sensitivity and wide spectral range.
|
Electronic and
micro-computer system
|
Automatic adjustment of light source power.
Light energy and negative high-voltage
auto-balance
|
Display mode
|
LED display of energy and measurement
values, concentration direct reading
|
Read mode
|
Transient, time average, peak height, peak area.
Integral time is selectable in the range of 0.1-19.9s.
|
Scale expansion
|
0.1~99
|
Data processing mode
|
Automatic calculation of mean, standard
deviation and relative standard deviation.
Repeating number is in the rang of 1-99
|
Measurement mode
|
Automatic curve fitting with 3~7 standards;
Sensitivity auto-correction
|
Result printing
|
Measurement data, working curve, signal
profile and analytical conditions can all be
printed out.
|
Instrument self-check
|
Check current status of each function key
|
Characteristic
Concentration and
Detection Limit
|
Air-C2H2 flame
|
Cu: Characteristic concentration≤0.025mg/L,
Detection limit≤0.006mg/L;
|
Function Expansion
|
Hydride vapor generator can be connected for hydride analysis.
|
Dimensions and
Weight
|
1020 x 490 x 540 mm, 80kg unpacked
|