The Frontier PPH Fume Hood provides the highest level of protection and containment against highly corrosive chemicals. Its full polypropylene interior and exterior makes it metal-free, therefore suitable for sensitive work such as trace metal analysis.
Key Benefits:
•Suitable for metal sensitive work
•Excellent aesthetics
•5 degree sloped front
•Unique perforated baffle design
•Superior containment at 0.5m/s
•High energy savings
Features
1. Full polypropylene(PP) interior and exterior
•Metal-free: suitable for sensitive work such as trace metal analysis
•Wall: stress-relieved, acid-resistant, non-corrosive, bend-resistant
•Work-top: excellent acid and alkali erosion endurance, high heat and decoloration resistance
2. Operates safely at 0.5m/s at 18" (457mm) or full open sash position constant volume concept eliminates the need forcomplex VAV systems.
3. Optional Sentinel XL Airflow Monitor accurately monitors face velocity and alarms when inflow decreases to unsafe levels.
4. The tapered PP exhaust collar reduces airflow turbulence, static pressure loss, noise level and increases face velocity uniformity.
5. Perforated Baffles: improve capture efficiency.
6. Air Foil: reduce turbulence and eliminate backflow.
7. Sturcture design
•Sash: tempered safety polycarbonate, can be fully opened and set in any working position
•Base Cabinet: one shelf fixed in the left cabinet to prevent liquid leakage
8. Remote control water fixture and 1 gas fixture are factory installed and plumbed, Up to 4 services may be factory or field installed.
9. Perfect Pitch Profile™: 5° Angled Front, this gentle pitch of the hood profile enhances visibility and maximizes operator comfort.
10. Removable structure for easy access: baffles, panel on internal side wall and rear access panel.
Models
Model
|
PPH-4UDP_W-_
|
PPH-5UDP_W-_
|
PPH-6UDP_W-_
|
Nominal Size
|
1.2 m (4')
|
1.5 m (5')
|
1.8 m (6')
|
External Dimensions (W x D x H)
|
1200x900x2350mm
47.2"x35.4"x92.5"
|
1500x900x2350mm
59.0"x35.4"x92.5"
|
1800x900x2350mm
70.8"x35.4"x92.5"
|
Internal Work Area (W x D x H)
|
980x665x1925mm
38.5"x26.1"x75.7"
|
1280x665x1925mm
50.3"x26.1"x75.7"
|
1580x665x1925mm
62.2"x26.1"x75.7"
|
Exhaust Outlet Diameter
|
250mm
|
300mm
|
300mm
|
Maximum Sash Opening
|
740mm (29.1’’)
|
Exhaust Volume/Static Pressure Required
at 0.5m/s (100fpm) at Full Open Sash
|
1305cmh @ 73Pa
768cfm @ 0.29”WG
|
1705cmh @ 95Pa
1004cfm @ 0.38”WG
|
2105cmh @ 117Pa
1239cfm @ 0.47”WG
|
Electrical Requirements
|
230V, AC, 50/60Hz
|
PPH-4UDPVW-8
PPH-4UDPCW-8
|
PPH-5UDPVW-8
PPH-5UDPCW-8
|
PPH-6UDPVW-8
PPH-6UDPCW-8
|
110-110-120V, AC, 50/60Hz120V, AC, 50/60Hz
|
PPH-4UDPVW-9
PPH-4UDPCW-9
|
PPH-5UDPVW-9
PPH-5UDPCW-9
|
PPH-6UDPVW-9
PPH-6UDPCW-9
|
Accessories
•Sentinel XL airflow monitor (discussed above).
•Additional service fixtures and electrical outlets.
•Exhaust blower
bio-equip.cn