1.Product Introduction
The HF-220 biochemistry analyzer is a small mobile fully automatic quality instrument, with multiple analysis method, ideal for various tests in small and medium-sized laboratories and clinical centers.
2.Technical Parameter
Reliable price fully automatic laboratory biochemistry analyzer instrument |
Working Mode |
Fully-auto, discrete, random access, measured in assay cup directly |
Testing Speed |
300 tests/h (single, double reagent) |
Colorimetric cup |
44 colorimetric cups , quartz cuvette is optional |
Stirring mechanism |
Independent Teflon coating stirring rod |
The cleaning of colorimetric cup |
3*4 steps automatic cleaning |
Sample position |
40 |
Reagent position |
80 reagent positions, (40 R1, 40 R2) |
Amount of reaction liquid |
150ul-350ul |
Reaction time |
10 minutes |
Temperature |
37°C ±0.1°C |
Injector |
ceramic trace plunger pump |
Measure Method |
end-point method, dynamic method, immunoturbidimetry, fixed time method, etc |
Measurable wavelength |
340-800 nm (single/dual wavelength) |
Wavelength |
340nm, 405nm, 450nm, 505nm, 546nm, 578nm, 630nm, 700nm ( four backups are optional) |
The range of absorbance |
0-4.0 Abs |
Resolution |
0.0001 ABS |
Sample volume |
2ul - 30ul, step by 0.1ul |
Reagent volume |
20-300ul, step by 1ul |
Operation system |
Windows XP, Windows 7, Windows 8 |
Interface |
USB |
Light source |
12 v long-life halogen lamp |
Power supply |
AC 220V, 50Hz, 1000VA |
Water consumption |
8L/h (max) |
Dimension |
650*700*1200(mm) |
3.Test Items
1.Liver function: ALT/GPT, AST/GOT, ALP, T.BIL, D.BIL, TP, ALB
2.Renal function: BUN, CRE, CO2, UA
3.Blood lipids: CHO, TG, HDL-C, LDL-C
4.Blood glucose: GLU
5.Myocardial enzymes: CK, CK-MB, LDH, a-HBDH, r-GT
6.Ion: Ca, P, Mg, K, Na, Cl
4.Operating Environment
Items |
Operating Environment |
Temperature |
15℃~30℃ |
Relative Humidity |
≤80% |
Table-board |
Please put the instrument on a stable, ventilate,
clean and flat table-board and with stable power
supply, no strong electromagnetic interference
|
bio-equip.cn