Home >Products> Medical Lab Instruments >Microplate Readers / Plates Washers> Microplate Reader
Medical Lab Instruments:  Microplate Readers / Plates Washers   Microbiological Analysis   Other Medical Lab Instruments  
Microplate Reader
Microplate Reader
Place of Origin:
United States
Brand:
REAGEN LLC
Model:
RNE90002
Supplier:
REAGEN kits
Price:
By e-mail
Hits:
1285 
Updated:
5/3/2020
  • Product Detail
  • Company Profile

    Weight:

    8.1kg

    Overall dimensions:

    430mm(L)×335mm(W)×175mm(H)

    Power:

    a.c.110V~220V,50-60Hz

    Fuses:

    T2AL250V

    Work Environment:

    temperature 10℃-30℃;Humidity ≤70%

    Store Environment:

    -20℃-55℃;Humidity ≤93%

    Standard Wavelength:

    405,450,492,630nm,Ten optical fibers can be configured within 400 to 800 at a time;

    Instrument sensitivity:            

    no less than 0.01L/mg.

    Instrument channel difference:

    less than 0.01A.

    Range of operating wavelength:

    400-800nm;

     

    Detection range:

    0.000-4.000A;

    Resolution ratio:

    0.001A;

    Plate reading velocity:

    5s of single waveform, 10s of dual wavelength;

    Warm up time:

    10 minute

    Interface:

    RS-232C serial interface, USB interface

    Display:

    5.7″LCD display(320×240 discernibility,256 gray scale)

    Input:

    Touch panel and pen, 

    Technical Specifications

    1) Wavelength characteristics: requirements on accurate wavelength and half wave width of optical fibers for instrument see the table below.

    Unit: nm

    Accurate wavelength

    Half wave width

    ±3.0

    ≤12

    2) Accurate absorbance and linear error: measurement accuracy and linear error of instrument shall comply with the provisions as described in the table below:

    Absorbance range

    Accuracy

    Linear error

    0.000-2.000

    ±0.01

    ±1.0%

    >2.000-3.000

    ±0.05

    ±2.0%

    >3.000-4.000

    ±0.15

    ±5.0%

    3) Repeatability of measurement of instrument: no greater than 0.5%.

    4) Instrument stability:

    a) Absorbance varies shall be no greater than ±0.005A within 10min due to drifting.

    b) If the power supply voltage of 220V varies by ±10%, the resulting variation of absorbance is no greater than ±0.005A.

    bio-equip.cn
    REAGEN LLC is an indrustry-leading biomedical enterprise focused on designing and producing a broad range of ELISA kits, including clinical diagnostic kits, food and feed safety kits, and veterinary disease kits. we also manufacture microplate readers, plate washers, and related instruments and equipment.
Request Information
Request Information: yes no
Request Quotation: yes no
refresh

I agree to share my inquiry with the other matching suppliers.

Request Information

* Name:
Job Title:
* Tel:
Fax:
* E-mail:
Postcode:
Institution/Company:
Address:
* Country:
Request Information:
yes no
Request Quotation:
yes no
* Message:
* Verification Code:
refresh
I agree to share my inquiry to the other matching suppliers.